Location:Home > Products > Agent Brand Instruments > Fuji-Japan & China

HPT604 Series Submersible Double Output Pressure Level Transmitters

1. Profiles:

HPT604 series digital and analog pressure level transmitter with high stable and reliable, which use GE silicon pressure sensor and high accurate circuit board into the stainless steel housing. Integrated construction and standard signal provide the user easy and convenient application in the local working place. The special cable connects with housing, can be immerged into the media for a long time. HPT604 double output designed incorporating with monolithic computer technology and sensor digital conversion technology, which core component adopts 24-bit AD MCU micro-processor to ensure high quality of the transmitter relaying on its strong function and high speed operation capacity. The overall designed framework is to meet the requirements of increasingly enhanced industrial site application with a view to reliability, stability, high accuracy and the product also features strong function and without manually operating device to ensure good interaction. Application digital signal processing technology is made for good disturbance immunity. 


2. Specifications:

Model:

Parameter:

HPT604 (Double Output Type)

Pressure Range:

0-0.1 Bar...50 Bar / 0-1M...500M LEVEL Optional

Programmable:

Zero adjustable by RS485 Communication.

Safe Overload:

150% F.S.

Burst Pressure

300% F.S.

Accuracy:

(Linearity Hysteresis

Repeatability)

 ±0.5%F.S;  ±0.25%F.S;  ±0.15%F.S

 ±0.1%F.S optional according to level range.

Including non-lin., rep. and hys

Long Stability:

Standard: 0.1%F.S±0.05%/Year

Max: 0.15%F.S±0.05%/Year

Working Temp:

-30~85

Storage Temp:

-30~100

Temperature Compensation:

-10~60

Medium compatible:

Compatible with 316L Stainless Steel or Titanium Alloy or PTFE   materials.

Output:

Double 4-20 mA

For Level &Temp

4-20mA + HART

For level

4-20 mA +Switch

(Relay output)

4-20mA + Temp Output (Ω)

RS485 for

 Level &Temp

Power Supply:

12~ 30 V DC (Standard)

Baud rate:

1200, 9600, 19200 optional (for RS485)

Data format:

1-bit start bit, 8-bit data bit, 1-bit stop bit (for RS485)

Odd and even calibration:

None

Sampling rate:

10 points/second to 1 point/50 minutes

Resolution:

20 bit 10ppm/ 0.1 mm

Loading Resistance:

(U-12)/0.02 Ω

Insulate resistance:

>100M Ω  @50V dc

Zero Temp. Drift:

Typical: 0.05%FS/

Max: 0.01%FS/

FS Temp. Drift:

Typical: 0.005%FS/

Max: 0.01%FS/

Electronic connection:

Fixed cable and water proof IP68

Response time:

10ms

Pressure Type:

Gauge pressure; Absolute pressure optional.

Certificate approving:

Exia IICT6 and CE Certificate.

EMC Standard:

electromagnetic radiation:EN50081-1/-2

electromagnetic susceptibility:EN50082-2

Lighting Proof:

Air conducts withstand voltage 8000V Shell and Cable conduct   withstand voltage 4000V.

Water Proof:

IP68

Temperature Range:

-10 to 100 Degree C optional

Temperature sensor type:

PT100

Switch type:

Relay(dry contact)

Weight:

Net weight is about 0.25KG/1KG, Full Packing weight is   about 0.5KG/3KG.

Remarks:

Special applications request by Customized


3.Part Number Selection Table:

604

Submersible   Double Signal Pressure Liquid Level Transmitters


code

Pressure & Level Range

code

Pressure & Level Range

00

0…0.5

21

0…7

01

0…1.0

22

0…8

02

0…1.1

23

0…10

03

0…1.2

24

0…12

04

0…1.3

25

0…15

05

0…1.4

26

0…16

06

0…1.5

27

0…20

07

0…1.6

28

0…25

08

0…1.7

29

0…30

09

0…1.8

30

0…35

10

0…1.9

31

0…40

11

0…2.0

32

0…50

12

0…2.1

33

0…60

13

0…2.2

34

0…80

14

0…2.3

35

0…100

15

0…2.4

36

0…150

16

0…2.5

37

0…200

17

0…3

38

0…250

18

0…4

39

0…300

19

0…5

40

0…500

20

0…6

X

By Customized


Code

Pressure & Level Units

H

m H2O  (Min: 0.5 mH2O; Max:500 mH2O )

B

Bar    (Min: 0.05Bar Max:   50Bar)

P

Psi     (Min:1Psi;   Max:725Psi)

F

m Fuel  (Min:0.5m; Max:500m)


Code

Pressure type

G

Gauge/Relative pressure type (universal)

A

Absolute pressure (customized)


Code

Power Supply

S3

24 V DC

S15

1230 V DC

X

By Customized


Code

Signal Output

E14

4-20 mA + HART Protocol 2 wires   (customized)

E11

RS485 (MODBUS) RTU for Level&Temp

E21

Double 4-20 mA for level and temp

E22

4-20 mA with temperature resistive signal

E23

4-20 mA with Switch signal (Relay output)

X

By Customized


Code

Measuring   Medium

CW

Common Water

SW

Sea Water

D1

Diesel, density 0.84

D2

Diesel, density 0.83

D3

Diesel, density 0.85

D4

Diesel, density 0.86

X

By Customized


Code

Others   Function (Optional)

N

Standard Type (with short filter)

TF

With Tapered Filter Covers

M1

G 1/4’’ male thread mounting type

NM

Nylon materials filter

TA

Titanium alloy materials (high cost!)

P

All Plastic Housing Structure


Code

Accuracy

1

≤±0.5%F.S

2

≤±0.25%F.S

4

≤±0.15%F.S (high cost!)


Code

Cable   Length

003

3 Meters

XXX

By Customized

604

01

H

G

S3

E21

CW

M1

2

002

Remarks

Example P/N60401HGS3E21CWM12002

sdf.jpgDatasheet



Contact Us
HOLYKELL TECHNOLOGY
Lugu High-Tech Development Zone
Changsha City,Hunan,China
 admin@holykell.com
 +86-0731-89873265
 +86-0731-89873646